Measurement of Cell Concentration in Suspension by Optical
Density
Scott Sutton,
Ph.D.
Vectech
Pharmaceutical Consultants
This article first appeared in the
PMF Newsletter
of August, 2006 and is protected by copyright to PMF. It
appears here with permission.
A common issue for the microbiology lab is the determination
of starting inoculum concentration. If the inoculum
concentration is determined by plating, the inoculum is several
days old before use. This essay describes the use of turbidity
to estimate microbial concentration in a suspension, using the
Antimicrobial Efficacy Test as the example.
Determination
of Inoculum for the AET
The compendial antimicrobial efficacy test
(AET) requires inoculation of the product with microorganisms to
a final concentration of approximately 106 CFU/mL.
Although this seems to be a minor point, it does serve to
illustrate some of the inherent difficulties in microbiological
testing and the need for experienced and academically trained
microbiologists to head the laboratory.
Let’s look at the compendial guidance. The
Pharm Eur (1) instruction on preparing the inoculum for the AET
states:
“To harvest the
… cultures, use a sterile suspending fluid … Add sufficient
suspending fluid to reduce the microbial count to about 108
micro-organisms per milliliter…Remove immediately a suitable
sample from each suspension and determine the number of
colony-forming units per milliliter in each suspension by plate
count or membrane filtration (2.6.12). This value serves to
determine the inoculum and the baseline to use in the test. The
suspensions shall be used immediately.”
There are, of course, two problems with
these instructions. The first is that the technician is
instructed to use an inoculum of about 108
microorganisms per milliliter and then instructed to determine
this by plate count. Colony forming units (CFU) and cells are
two different measures and this will inevitably lead to
difficulties as the unfortunate lab worker cannot guarantee the
number of cells in the suspension, only the number of CFU
found. However, we can accept the scientific inaccuracy as the
numbers will generally work out. The more serious problem is
the instruction to use the plate count CFU for determination of
the inoculum for the test, and that the suspension shall be used
immediately. This quite frankly cannot be done. If you use the
suspension immediately, the plate counts are unavailable, if you
use the plate counts to set the inoculum, then the suspension is
at least a day old.
Contrast these instructions with those in
the USP (2) for the same exercise:
“To harvest the
… cultures, use sterile saline … Add sufficient … to obtain a
microbial count of about 1 x 108 cfu per mL…[Note:
The estimate of inoculum concentration may be performed by
turbidimetric measurements for the challenge organisms.
Refrigerate the suspension if it is not used within 2 hours].
Determine the
number of cfu per mL in each suspension …to confirm the initial
cfu per mL estimate. This value serves to calibrate the size of
the inoculum used in the test.”
These USP instructions have the advantage
of being physically possible to perform, an obvious advantage to
the lab worker. However, the turbidometric measure of the cells
is also only an approximation of CFU. Thus the instruction to
confirm the numbers (after the test is underway) with the plate
count is an important control on the test.
This article will explore the turbidometric
approximation for cell numbers, and important controls on the
process as well as potential pitfalls to the method.
Theory
Light
scattering techniques to monitor the concentration of pure
cultures have the enormous advantages of being rapid and
nondestructive. However, they do not measure cell numbers nor
do they measure CFU. Light scattering is most closely related
to the dry weight of the cells (3).
Light is passed through the suspension of
microorganisms, and all light that is not absorbed is
re-radiated. There is a significant amount of physics involved
in this, and those interested are referred to optical treatises,
particularly those discussing Huygens’ Principle (a good choice
is
Light Scattering by Small Particles by H C Van De Hulst).
For our purposes it is enough to say that light passing through
a suspension of microorganisms is scattered, and the amount of
scatter is an indication of the biomass present in the
suspension. In visible light, this appears “milky” or “cloudy”
to the eye (3). It follows from this that if the concentration
of scattering particles becomes high, then multiple scattering
events become possible.
Methods
McFarland Turbidity Standards
McFarland standards can be used to visually
approximate the concentration of cells in a suspension. The
McFarland Scale represents specific concentrations of CFU/mL and
is designed to be used for estimating concentrations of gram
negative bacteria such as E.
coli. Note that this estimate becomes uncertain with
organisms outside the normal usage as different species of
bacteria differ in size and mass, as do yeast and mold. Use of
this method would require calibration and validation.
McFarland Standards are generally labeled
0.5 through 10 and filled with suspensions of Barium salts.
(Note - latex bead suspensions are now also available which
extend the shelf life of the material). The standards may be
made in the lab by preparing a 1% solution of anhydrous BaCl2
and a 1% solution of H2SO4 – mix them in
the proportions listed in the table. They should be stored in
the dark, in a tightly sealed container at 20-25oC
and should be stable for approximately 6 months (4).
The advantage of the use of these standards
is that no incubation time or equipment is needed to estimate
bacterial numbers. The disadvantage is that there is some
subjectivity involved in interpreting the turbidity, and that
the numbers are valid only for those microorganisms similar to
E. coli. In addition,
the values are not in the appropriate range for the AET inoculum
and so further dilutions may be required.
Approximate
E. coli
concentrations on McFarland Scale
|
McFarland Scale |
CFU (x106/mL) |
1% BaCl2/
1% H2SO4
(mL) |
|
0.5 |
<300 |
0.05/9.95 |
|
1 |
300 |
0.1/9.9 |
|
2 |
600 |
0.2/9.8 |
|
3 |
900 |
0.3/9.7 |
|
4 |
1200 |
0.4/9.6 |
|
5 |
1500 |
0.5/9.5 |
|
6 |
1800 |
0.6/9.4 |
|
7 |
2100 |
0.7/9.3 |
|
8 |
2400 |
0.8/9.2 |
|
9 |
2700 |
0.9/9.1 |
|
10 |
3000 |
1.0/9.0 |
Spectrophotometer
The spectrophotometer
method measures turbidity directly. The best case (i.e.
most sensitive) would be to have a narrow slit and a
small detector so that only the light scattered in the forward
direction would be seen by the detector. This instrument would
give larger apparent absorption readings than other instruments.
As should be obvious, each
spectrophotometer used must be independently calibrated for use
in estimating microbial concentrations. Not only is the
apparent absorption affected by the width of the instrument’s
slit, the condition of the filter, and the size and condition of
the detector, but also each time the lamp is changed the
calibration needs to be repeated as different bulbs may vary in
total output.
The correlation of absorption to dry weight
is very good for dilute suspensions of bacteria (5), and this
relationship seems to hold regardless of cell size (although the
relationship of absorption to CFU does not). However, in more
concentrated suspensions this correlation (absorption to dry
weight) no longer holds. The linear range of absorption to
estimated CFU is of limited scope and for this reason the
calibration study must demonstrate the linear range of the
absorbance vs CFU values
and the relevant values.
Procedure
As there are a variety of different
instruments, there cannot be one single procedure. In general,
the spectrophotometer can be set at a wavelength of 420 – 660
nm. This wavelength must
be standardized and may need to be adjusted specifically to the
material being tested. Different vegetative cells, bacterial
spores and spores of
Aspergillus niger may not have the same maximal
absorbance wavelength.
It is important to have the cells in known
physiological state of growth. That is to say, as the cell size
varies with phase of growth (lag, log, stationery) the
approximate relationship between absorbance and CFU will also
vary. A recommended practice might be to pass a single
well-isolated colony twice on overnight cultures surface streaks
from the refrigerated stock, harvesting the rapidly growing
culture from the second passage for preparation of vegetative
cells. This also will serve to minimize a source of variability
for the AET (6).
A second source of concern might be the
cuvette used for the measurement – care must be taken to
maintain the correct orientation of the cuvette, and to protect
it from damage that could affect the passage of light. Finally,
it is necessary to blank the spectrophotometer (adjust the
absorbance reading to zero) using a standard, either water or
the suspending fluid, and maintain this practice.
Calibration
It must be stressed that this calibration
should be done for all organisms. The size of the organism, any
associated pigments, the preparation of the suspension, and
other factors all influence the readings. This calibration
study should also be rechecked after changing the bulb on the
light source, and should be reevaluated throughout the life of
the light bulb.
The calibration itself is simple to
perform. Prepare a concentrated solution of the organism, grown
under the conditions that will be used for the test. Make a
series of dilutions to cover the range of absorption
measurements of interest; 5 to 8 dilutions are recommended.
Immediately take the spectrophotometer readings in sequence, and
then take a confirmatory reading of the first in series to
confirm that no growth has occurred. The dilutions are then
immediately plated for viable count (serial dilution of the
suspensions will be necessary). Graph the relationship between
the absorbance and the CFU/mL after the plate counts are
available and use values in the linear range of this graph.
As there are several factors that can
affect this curve (quality of lamp output, size of slit,
condition of filter, condition of detector, microorganism
characteristic, etc)
this calibration should be confirmed when the conditions of the
assay change.
Conclusions
The use of optical density to estimate CFU
in a suspension is possible, if basic precautions are taken. It
is important to control:
-
The physiological state of the organism
-
The species of the organisms (i.e.
don’t calibrate the instrument using
E. coli and expect
the numbers to work for
Candida albicans)
-
The nature and condition of the equipment
Despite the inherent inaccuracy of the
method, if the procedure is adequately controlled and calibrated
the estimation of microbial numbers by optical density (either
by McFarland Standards or spectrophotometrically) is
sufficiently accurate for use in preparing inocula for QC
testing and offers the overwhelming advantages of being rapid,
low cost and non-destructive.
References
- EP. 2006. 5.1.3 Efficacy of Antimicrobial Preservation.
Pharm Eur. 5.0:447-449.
- USP. 2006. <51> Antimicrobial Effectiveness Testing
United States Pharmacopeia 29:2499-2500
- Koch, AL. 1994. “Growth Measurement” IN:
Methods for General and Molecular Bacteriology Gerhardt,
P et al (ed) American Society for Microbiology, Washington,
DC. p. 248-277.
- Smibert, RM and NR Kreig. 1994 “Phenotypic
Characterization” Section 25.4.9 IN: Methods for
General and Molecular Bacteriology Gerhardt, P et al (ed)
American Society for Microbiology, Washington, DC. p.
607-654.
- Koch, AL. 1970. Turbidity Measurements of Bacterial
Cultures in Some Available Commercial Instruments. Anal
Biochem 38:252-259
- Gilbert, P. et al. 1987. Inocula for Antimicrobial
Sensitivity Testing: a Critical Review. J Antimicrob
Chemother. 20:147-154.
Consulting with Scott Sutton |